Features |
- Color TFT Display,English/Chinese Manual,4 Waveforms
- Designed Weight-lighter to take, Built-In Battery,Continuously Monitoring
- Parameters:ECG,Respiration,NIBP,1-Temp,SPO2
- Abnormal HB,ST Segment Ayalysis
- Maximum 72 Hours Graphic and Tabular Trends Of All Parameters with Audioable And Visual Alarm
- Built-in Recorder with numerous print Formats Selectable
|
Specifications |
| Safety |
| BM-9000S colorful and easy-taken multi-parameters monitor has the same safty standards which agrees with the standards of IEC60601 Groups and GB9706 |
| Dimension and Weight |
Dimension: 318mm (W)×152mm(D)×264mm(H)
Weight: 5.4kgs |
| Operation Environments |
Power Requirements: 100-250V/AC, 50/60 Hz
Application Scope: Neonatal, pediatric, and adult patients |
| Performance Specifications |
- Display: Colorful TFT display
- Pixel: 640×480
- Language: English
- Indicator:
4-waveform;
Alarm Indicator;
AC Power Indicator;
QRS beep and Alarm Sound
- Interface:
Parameter module interface;
AC Power input socket;
External VGA Two display mode;
simulative signals output
- Battery: Rechargable, supplying power more than 2 hours
- Trend time: 1~72 hours
- Alarm:
User-adjustable high limit and low limit,
3 Level Audible and Visual Alarming
- Recorder: Built-in thermal printing system
- Plethysmogram waveform: 2 channels
- Recording modes: manual,alarm,timing
- Print speed: 25mm/s, 50mm/s
- Network: Connection to center monitor system
|
| ECG Specifications |
- 3-lead
- Lead Input: RA; LA; LL
- Lead mode: I; II; III
- Scanning Speed: 12.5 mm/s; 25mm/s; 50mm/s
- Gain selection: ×0.25; ×0.5; ×1; ×2; auto
- Heart Beating Rate:
Adult: 15-300bpm(B/S), Pediatric/Neonate: 15~350bpm
- Accuracy: ±1bpm
- Resolve Rate: 1bpm
- Leakage: Less than 3μA
- Protection:
Isolation of Voltage AC4000V/50Hz,
Resistance Against Defibrillator/Electronic
knives and skin voltage
- Scaling signal: 1mV/±3%
- Bandwidth: 0.05~130Hz
- Alarm range: 15~350bpm
- ST detection: YES
- Measurement range: -2.0mV~2.0mV
- Alarm range: -2.0mV~2.0mV
- Analysis & Recall of Abnormal heart beating: YES
- Alarming: Audible and visual alarms/recall
|
| Respiration Specifications |
- Mode: RA-LL impedance
- Range: Adult: 0~120rpm;
- Pediatric/Neonate: 0~150rpm
- Resolve Rate: 1rpm
- Accuracy: ±1rpm
- Alarming range: 0~150rpm
- Apnea alarming: YES
|
| NIBP Specifications |
- Measurement Method: Auto surge mode
- Working model: Manual/Automatic/Continuous
- Auto measure time: selectable
- Measurement unit: mmHg/Kpa selectable
- Alarm: Systolic, Diastolic, Mean
- Measurement range:
Adult: 10~270mmHg;
Child: 10~200mmHg;
Neonate: 10~150mmHg
- Over-pressure protection: yes
- Resolution rate: 1bpm
- Accuracy: ±1rpm
- Pulse Alarm: 0~254bpm
|
| Temperature Specifications |
- Measurement range: 0~50°C
- Resolution: 0.1°C
- Accuracy: ± 0.1°C
- Alarm range: 0~50°C
- Channel: one channel
|
| SpO2 Specifications |
- Measurement range: 0~100%
- Resolution: 1%
- Accuracy: ±2% (70~100%); ±3%(40~69%)
- Alarm range: 0~100%
- Pulse Rate: Range: 0~250bpm
- Alarming Range: 0~100%
- Pulse Range: 0~254bpm
- Resolution: 1bpm
- Accuracy: ±1bpm
- Pulse Alarm: 0~254bpm
|
| Net weight : 4.0 KGS |
| Gross weight : 8.2 KGS |
| Configuration : ECG+Breathe+NIBP+SpO2+HR+1-TEMP+Battery(one piece) |
| Optional : Recorder |