Features |
| Anesthesia apparatus |
- Equipped with 2 vaporizers namely Enflurane and Isoflurane, which can be used alternatively vaporizer has the function of temperature control, flow and pressure compensation with an induction-lamp. and easy to adjust.
-
Flow meter range O2-N2O interlock with high precision O2:0.1~10L/min, N2O:0.1-10L/min, Air :0 1-10L/min
- Integrated circuit system: power cutout protection device. UPS rechargeable battery back up 2 hours after mains supply having been cut.
|
| Ventilator |
- Gas driven electrically controlled, lime switch, volume constant and pressure restriction.
- Monitor large screen LCD display with high brightness and wide visual angle, microcomputer control, multifunctional respiration monitor, monitoring tidal volume. Ventilatory volume, gas-feeding pressure, airway pressure, respiratory frequency. P-T and P-V waves etc
|
Safety Device |
- Alarm upon power having been cut. N2O is cut off automatically if O: is insufficient; alarm upon the excess of upper & lower airway pressure limit and ventilatory volume.
|
| It can be assembled with ECG monitor and Monitor for O: concentration. |
Optional: UPS. |
| Optional: Pen Ion vaporizer or Perlong vaporizer. |
Specificatiion |
| Suitable for: |
Adult, pediatrics |
| Driving Method: |
Gas driven electrically controlled |
| Ventilation modes: |
IPPV, SIPPV, SIMV, PEEP |
| Tidal Volume: |
50-1200ml |
| Max Insp: |
Flow 18L |
| Respiratory Rate: |
-1~7.0Pa(=5%) |
| I:E: |
1:0.5,1:1, 1:1.5, 1:2.0,1:2.5.1:3.0 |
| Max pressure: |
< 6kPa |
| PEEP: |
0.1~1.2KPa |
| Sigh Volume: |
not less than 1.5 times of Tidal Volume |
| Upper airway pressure limit: |
1.0kPa ~ 5.4kPa |
| Lower airway pressure limit: |
≥ 0.5kPa |
| |
Continuous work after mains interruption |
| O2Supply Speed: |
35L/min~75L/min |
| |
 |
|
|