Features |
| High resolution thermal printing array system
|
| High frequency response reach to 150Hz
|
| Capable of printing continuous one channel trace and annotations including lead mark
and such parameters as sensitivity, paper speed, and filter operation status. |
| High resolutioin digital filter inhibits baseline drift without distorting ECG wave-form.
|
| One-touch operation able to enhance efficiency and relieve operating strain.
|
| Concept of floating input circuit meets IEC safety standard.
|
| Visual status due to light indicated touch-keys.
|
| Rechargeable battery supporting more than 50 patient examinations.
|
| Under the automatic mode, pressing key “MODE” continuously can keep recording
when abnormal lead is found. |
Specifications |
| leads |
Standard 12 leads manual/auto |
Input Circuit |
Protect instrument from tilting and possible vibration or shock |
Input Impedance |
>50MΩ |
| Output Impedance |
<100Ω |
Patient current leakage |
<10μA |
| Input Circuit Current |
≤50 nA |
| Calibrating Voltage |
1mV±5% |
| A/D Conversion |
13 bit |
Frequency Response |
0.05~150Hz |
| Time Constant |
>3.2s |
| CMRR |
>80dB
>100dB(with AC filter) |
| EMG Filter |
35Hz(-3dB)/25Hz(-3dB) |
| Sensitivity Setting |
5、10、20mm/mV,conversion tolerance±5% |
| Noisy Level |
≤15μVp-p |
Power supply |
AC: 220/110V, 50/60Hz,15VA
DC: 12V rechargeable battery |
| Safety standard |
IEC I/CF with protecting circuit for vibration or shock |
| Recording System |
Thermal-array printing paper roll 50mm×20m |
Depolarization Voltage |
±300mV |
Baseline Control |
Auto-adjustable |
| Nosie Level |
≤15μVp-p |
Dimensions |
(L×W×H) 310mm×230mm×70 mm |
| Dimension of package |
(L×W×H) 410mm×280mm×200mm |
Net weight |
2.69 KGS |
Gross weight |
4.65 KGS |
 |