General Features |
| Application of high precision thermal array printer to ensure clear ECG waveforms and rich literal interpretations for clinical diagnostic and research purpose. |
| Application of high speed thermal paper 50mm width to record ECG waveforms, measured parameters and optional diagnostic reports. |
| Synchronously acquisition of 12-lead ECG waveforms provides reliable data for diagnostic purpose. |
| Application of digital isolation technology ensures safety and reliability of the device. The influence of temperature/time drift is reduced to the lowest and guarantees the adaptability to environment. |
| Digital signal processing of drift control, AC filter, EMG filter and heart measurement ensures the reality and reliability of signal processing. |
| Auto measurement and optional diagnosis of conventional ECG parameters doctor’s burden. |
| Three printing formats, satisfy any kinds of clinical requirements. |
| LCD Modual display Menu/Status. |
| Selectable II and V5 rhythm lead makes it convenient to observe abnormal heart beat rate. |
| Automatic adjustment of baseline drift controls effectively baseline drift, optimizes printing position and ensures high quality ECG waveforms. |
| AC and DC power supply, built-in rechargeable battery, recharging circuit, and perfect battery management and protection system. |
| Strictly follows IEC60601-1 ClassⅠ, type CF safety standard. |
| Compact and delicate design, friendly menu makes it easy to operate and releases operation strain. |
Special Features |
| With RS232 interface, can be upgraded with software to connect with PC.(optional:12-lead ECG waveform can be transferred to PC, measured, analyzed, stored or printed out on A4 paper) |
| High quality measurement with precise and consistency analytic results. |
| ARR analytic report results. |
Specifications |
| Input Circuit Floating |
floating, Protection against defibrillator effect |
| Lead Standard |
12 leads |
| Patient Current Leakage |
<10μA |
| Input Impedance |
≥50ΜΩ |
| Calibrating Voltage |
1mV±3% |
| Drift filter |
Anti-drift filter |
| Filter |
AC,EMG filter |
| Digital Sampling |
18 bits/1000Hz (simultaneously acquisition of 12 leads ECG data) |
| Frequency Response |
0.05Hz~150Hz |
| Time Constant |
>3.2seconds |
| CMRR |
>60dB |
| |
>100dB (with AC filter) |
| Sensitivity |
AUTO2.5, 5, 10, 20mm/mV |
| Recorder |
High resolution thermal printer |
| Paper Speed |
25, 50mm/s, ±3% |
| Input Circuit Current |
≤0.1uA |
| Power Supply |
AC 110V/220V/85v-265v / 50/60Hz ;30VA |
| |
DC 12V, 1500mA (Built in rechargeable battery) |
| Noise Level |
<15μVp-p |
| Threshold |
≤20uV |
| LCD Display |
Display menu,operation status and ECG waveform |
| Dimension |
288 x 204x 60mm |
| Dimension (carton) |
380 x300x200mm |
| Net weight |
2.75KGs |
| Gross Weight |
4.5KGs |
|